Advanced Laser Solutions Ltd.
 
 
  UM110 - Specification  
Units available:   Typical max. power, beam diameter dependent.
  UM110-50
UM110-50l
UM110-50d
CO2 6kW
CO2 3kW (improved low power performance)
CO2 & Nd:YAG 4kW
   
 

Key features:

The UM110 works on a sample on demand principle.
The device is inactive until a sample is requested, a sample then taking a fraction of a second.

UM110 contains inbuilt analogue and digital averaging allowing large amounts of data to be processed in a short time - typically 26 samples in a second. This allows rapid analysis of unstable, RF powered or pulsed beams.

 
     
   

Details:

Monitor any unfocussed CW laser, or pulsed with averaging (sensor and needle dependent).

Clear aperture 50mm

Power range From mW to kW (sensor and needle dependent).
 
Maximum sample averaging rate 26Hz,   Maximum data rate 4Hz, limited by RS232 serial and operating system.

Sample on demand or on timer.

Typical absorption < 0.24%.
 
Interface RS232.

PC (Win98/2000/NT/XP) or PDA based user interface.
 
Measurements X and Y axis intensity profiles.
Beam centre position.
Beam diameter.

Compact Housing System mount:, 180mm dia x 90mm.

Power requirements
Line frequency: 50-60Hz.
Power consumption: max 1.5A, typically 0.9A.
 
Weight
Net: approximately 1.5 kg
Shipping: approximately 4.5 kg

 
 
Please contact AdLas for further specification and demo software.